To view the performance of the display after calibration and conformance through easily understandable graphs, click on the view history option in the main window.
The history window appears as shown, you can see the information like task name, date and time of the task performed, serial number of the calibrated display and like.
Double click on the task name you wish to see the results.
The software displays graphical results for the corresponding tasks. Click the task names below to view various graphs based on the performed tasks.
For GSDF Color Caibration, Five graphs are generated. they are as listed below.
1.Luminance Response
2.Gamma Correction Curve
3.JND Conformance
4.AAPM Conformance
5.DeltaE
For GSDF DICOM calibration , Four graphs are generated which are listed below.
1.Luminance Response
2.Gamma Correction Curve
3.JND Conformance
4.AAPM Conformance
For DICOM Conformance, there are three graphs generated
1.Luminance Response.
2.JND Conformance.
3.AAPM Conformance.
Consistency Graph : option in the software history is very useful in tracking the consistency of your display overtime.
Select the display name and then provide the 'from' and 'to' date for which you wish to check the display's consistency performance.
Graphs are shown for black level consistency and white level consistency.